The dry biochemical analyzer is a portable dry biochemical quantitative analysis instrument. The analyzer uses the principle of reflection spectrophotometry to quantitative detect the clinical chemical components in the whole blood, serum and plasma of human blood samples by testing the matched test card that manufactured by Jiangsu Konsung Bio-Medical Science And Technology Co., Ltd.
Details:
The dry biochemical analyzer is a portable dry biochemical quantitative analysis instrument. The analyzer uses the principle of reflection spectrophotometry to quantitative detect the clinical chemical components in the whole blood, serum and plasma of human blood samples by testing the matched test card that manufactured by Jiangsu Konsung Bio-Medical Science And Technology Co., Ltd. The working principle of the dry biochemical analyzer is as follows: illuminate the color area of test card with a specific wavelength LED light source, measure the absorbance of the test card, capture the reflected signal strength with the photosensitive diode, and detect the content of the analyte according to the obtained AD value.
Parameters:
Electric shock protection | Class II equipment and internal powered equipment |
Degree of protection against liquid | IPX0 |
Operation mode | Continuous |
Product life | Three years (20 samples per day) or 22,000 sample counts |
Operating environment | Temperature: 15°C~35°C; Relative humidity: ≤85%RH |
Operating atmospheric pressure | 70kPa~110kPa |
Storage and transportation environment | Temperature: -20°C~60°C; Relative humidity: 10%~90% |
Barometric pressure | 50kPa~110kPa |
Measurement method | Photometric detection method |
Test Items | TC(Total cholesterol), TG(Triglyceride ), HDL-C(High Density Lipoprotein Cholesterol), LDL-C(Low Density Lipoprotein Cholesterol), GLU(Glucose) |
Specimen Type | Serum, plasma, whole blood |
Test time | ≤3min |
Precision | relative offset ≤±10% |
Precision (glucose) | In the concentration range of 5mmol/L~22mmol/L, the coefficient of variation is (CV,%) ≤5.0% |
Linear (glucose) | In the linear range of 1.1mmol/L to 33.3mmol/L, the slope is 1.00±0.05, and the correlation coefficient is r≥0.975 |
Stability | relative bias ≤ ± 10%. |
Data Storage | 3000 results can be stored |
Display screen | 4.3 inch touch screen |
Power supply | Adapter, input: AC100~240V, 50/60Hz, 30VA; output: 5V, 2A;Internal power supply:DC3.7±0.3V |